Basic functions
High-voltage circuit
Principle of operation
The high-voltage circuit is to reach the selected high-voltage as quickly as possible and
keep it constant during the exposure.
The high-voltage circuit is made up of the following components:
- D916 (CPU)
- Main inverter on D962
- Oscillating circuit capacitor C1
- High-voltage transformer in the single tank
- D907 measurement data log
- Intermediate circuit, capacitor bank D972
The high-voltage inverter is not active in standby.
With "exposure preparation", the nominal kV of D916 is output (measurable at TP-KVS
D916).
During exposure, D916 controls and monitors the high-voltage.
The high-voltage can be measured as test points TP-KVN, TP-KVP (1 V = 10 kV) and
TP-KV (1 V = 30 kV, total kV) on D916.
•
Possible error messages of the high-voltage circuit:
Sample high-voltage measurements
Positive/negative high-voltage measurement
•
Switch off the system
•
Remove the top cover and, in the case of MOBILETT XP Digital, place it on its side on a
table or similar surface (do not disconnect any cable connections)
Siemens AG
Medical Solutions
ERR 4
Static short-circuit at the main inverter
ERR 8
Static kV measured value <> 0
ERR 31 kV exceeds permissible limit
ERR 34 X-ray current exceeds permissible limit (tube short-circuit)
ERR 35 Short-circuit at the main inverter
ERR 37 Error in the kV control loop
SPR8-230.840.01.05.02
11.05
Page 57 of 72
CS SD 24
57
0
0
0
MOBILETT XP