Siemens MOBILETT XP Troubleshooting Manual page 9

Hide thumbs Also See for MOBILETT XP:
Table of Contents

Advertisement

General information
DANGER
Switch the unit off before servicing or maintenance. Always disconnect the power plug
first.
XP Hybrid: Make sure that the main switch and the battery mode are switched off.
The capacitor bank discharges to < 40 V in approx. 15 minutes.
Prior to performing any work, it must be verified that areas with dangerous voltage are
voltage-free
Siemens AG
Medical Solutions
To avoid electrical shock from components under voltage, also be
aware that:
The capacitors of the capacitor bank can be electrically charged
even when the system is switched off and the power cable is dis-
connected!
In the case of an error, individual capacitors of the capacitor bank
can still be electrically charged when disconnected from the
charging circuit!
Disconnect the battery blocks in the XP Hybrid/Digital prior to
maintenance and service work!
Carelessness can result in death or serious bodily injury.
Make sure that no parts or tools fall into the unit;
Do not touch potentially dangerous
components(Fig. 2 / p. 10)
If loose parts must be removed from the unit, use only
insulated tools;
Protect the work area so that no other persons are able
touch the unit while the covers are open or removed!
The safety covers can be removed after this period has elapsed.
(Fig. 2 / p.
10).
SPR8-230.840.01.05.02
11.05
Page 9 of 72
CS SD 24
9
MOBILETT XP

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mobilett xp ecoMobilett xp hybridMobilett xp digital

Table of Contents