Product-specific safety measures

DANGER
DANGER
DANGER
MOBILETT XP
While performing maintenance and service work
on the MOBILETT XP/Eco/Hybrid/Digital with the covers removed,
it is possible to come into contact with components under voltage.
Carelessness can result in death or serious bodily injury.
When performing maintenance and service work, the following is
to be observed:
the product-specific safety information contained in the
technical documentation,
and the general safety information (TD00-000.860.01...).
Remove or install components only if:
- The system is switched off, and
- The capacitors are discharged, and
- The ESD guidelines are followed, and
- The batteries have been disconnected.
Releasing radiation:
Checks and settings for which radiation must be
released are to be marked with the radiation warning
symbol.
Radiation protection measures are to be used.
SPR8-230.840.01.05.02