Error messages
ERR 36
System
mode:
Exposure
ERR 37
System
mode:
Exposure
ERR 38
System
mode:
Exposure
Siemens AG
Medical Solutions
Short-circuit at the filament inverter
Explanation: D916 (CPU) detects a short-circuit at the filament inverter
during an exposure.
Action:
Check test point TP - SCH on D916 (CPU) (high = error, short-circuit is
detected).
Check the maximum heating rate at TP - FC1, D916 (CPU)
Check the connections between D927 (X5) and the single tank (termi-
nals 11 and 12).
Possible defects: D927, D916.
Error in the kV control loop
Explanation: D916 (CPU) detects an error in the kV control loop.
Action:
The value KV
actual
(TP-KVS).
Check the cable connections between D962, C1 and the single tank,
and between D916 (X8) and the single, D907.
Check the capacitors of the capacitor bank, all LEDs must light on D972
when the MOBILETT XP is "on".
Possible defects: Fuse F1..F12, F14 (80A) on D972; C1..C12 D972, D962;
D907; D916; C1 resonance capacitor.
Error in the mA control loop
Declaration: D916 (CPU) detects an error in the mA control loop (test point
D916 (TP - JR) and (TP - JRS).
Action:
Check the maximum frequency of filament inverter FC1.
Ensure that value I
Check the cable connections between D916 (X8) and single tank D907.
Possible defects: D916; D927; D907.
SPR8-230.840.01.05.02
11.05
(TP-KV) does not agree with the value KV
(JR) agrees with value I
actual
Page 43 of 72
CS SD 24
43
nominal
(JRS).
target