Philips IntelliVue MX40 Instructions For Use Manual page 25

Reusable adapters and patient cables
Hide thumbs Also See for IntelliVue MX40:
Table of Contents

Advertisement

Inspect Cable Connector Gasket
Before connecting a cable to the MX40 monitor, visually ensure that the gasket
surrounding the mouth of the cable connector is seated properly (reference the
graphic at the front of this document that shows proper gasket seating). If any part of
the gasket is dislodged or damaged DO NOT USE! Instead, discard the cable as
there is a potential for incorrect readings and/or patient/user injury.
Inspect Cable Connector Contacts
Note: A cable's SpO
protective cap when not in use.
The cable's monitor connector has very small contacts that must be kept clean to
establish good connection between cable and monitor. Therefore, before connecting
a cable to the monitor, visually confirm there is no lint, moisture, or residue inside
the cable connector. If the visual check indicates a need to clean inside the cable
connector follow the steps below.
Important! Use the following steps only to clean inside the cable MX40 connector.
To clean connector or cable external surfaces, reference the validated cable cleaning
procedures provided in the Cables and Lead Sets Care, Cleaning and Disinfection
document that accompanies this IFU.
To clean inside the cable's MX40 connector:
1.
Prevent cleaning fluid from seeping into the cable connector and be careful not to dislodge
the gasket that surrounds the inside of the connector.
2.
Keep the cable connector opening pointed down as shown in the Clean Connector Pins
graphic at the front of this IFU; i.e. this helps prevent cleaner residue from building-up
inside the connector.
3.
Use only Isopropyl alcohol on a modestly dampened lint-free cloth, or pre-moistened
Isopropyl alcohol wipe.
4.
Gently wipe inside the cable's MX40 connector using very light swipes.
5.
Allow the connector pins to air dry prior to use.
CONNECTING AND DISCONNECTING THE MX40
Connecting: Use a "straight-on" pushing motion between the cable and MX40
connectors as shown in the Connect Cable graphic at the front of this IFU. A slight
clicking sound signifies a secure connection.
Disconnecting: Grip the sides of the cable connector and use a slight pivoting or
rocking motion to separate the connector from the MX40 as shown in the
Disconnect Cable graphic.
Caution: To avoid damaging a cable, NEVER disconnect by holding or pulling
on the cable connector wires as shown in the Never Pull Wires graphic.
PATIENT APPLICATION
Refer to the MX40 monitor IFU for information regarding proper ECG
lead/electrode placement that complies with standard AAMI or IEC practices.
Refer to the appropriate SpO
sensor being connected to a ECG/SpO
sensor connector should remain covered with its tethered
2
sensor IFU for proper patient application of any SpO
2
adapter cable.