Fig. 5
Replacing the NVRAM
Install the NVRAM on the CPU board (refer to Fig. 5).
Install the frame buffer board.
Establish all cable connections and ensure that they match their original status.
After having installed a new NVRAM, ensure that the system can function with this "new"
CPU. The following three additional steps are necessary:
Change the parameter values of the NVRAM to the correct values.
Replace the old Internet address with the new one.
Replace the old CPU no. with the new one.
MAMMOMAT 3000 Opdima
Service
Register 5
SPB7-230.061.02
Rev. 02
Page 6 of 20
05.99
SPS-UD
NVRAM





'
3
2
2
Siemens-Elema AB
Solna, Sweden