Follow all general safety guidelines when replacing the Camera Interface D100.
Note the special safety guidelines indicated in these instructions.
Fig. 11 Card retainers
Siemens-Elema AB
Solna, Sweden
Register 5
SPB7-230.061.02
Rev. 02
Open the top cover of the box (refer
to chapter 1).
Locate the Camera Interface D100
(refer to Fig. 10).
If there is a PCMCIA board
installed, remove the PCMCIA
interface (refer to chapter
"Removing the PCMCIA Board").
Open the SBUS card retainers.
Lift the card from its position.
Carefully lift the corners of the D100
to release it from the connector



Remove the board and place it on


'
the grounded tabletop pad.