Time - ABB SM500F User Manual

Field mountable paperless recorder
Hide thumbs Also See for SM500F:
Table of Contents

Advertisement

SM500F
Field mountable paperless recorder

7.6.3 Time

76
Set the date and time formats and set the current date and time using the
dialog box.
Note.
If daylight saving is required, enter the settings (see next page)
setting the time and date as the operation of the internal
before
clock is affected by the daylight saving settings.
The date and time cannot be adjusted if recording is enabled
during configuration, i.e the 'Disable recording in configuration'
check box is
not
Fig. 7.2, page 67.
Changes to the date and time are effective immediately upon
selecting 'OK' in the dialog box. Select 'Cancel' in the dialog box to
exit date and time setup without saving changes. Selecting
'Cancel' upon exiting Configuration Mode (see Fig. 7.7, page 73)
reset the clock to its previous setting.
does not
Setting an earlier date or time results in the loss of all data currently
in the internal buffer memory past that date. Data archived to
external media is unaffected. If an earlier time must be set, change
the Instrument Tag (see page 74). This causes new archive files to
be created and the duplicated hour of data is then saved to the
new files.
Time changes due to automatic daylight saving do not affect the
recorded data.
Archive files created during the daylight saving period (see
Section 7.7.4, page 90) are compatible with the database feature
of Version 6.2 (or later) only of the Company's DataManager data
analysis software package.
If the
status icon is displayed the clock battery must be
replaced – contact factory.
ticked on entry to the Configuration level – see
7 Configuration
IM/SM500F Rev. Z

Advertisement

Table of Contents
loading

Table of Contents