Figure 5-6. Wiping the wash stations with lint-free towels
6. Reinstall the wash head:
a. Replace the wash head using the alignment pins located on either side of the
retaining screw, then tighten the 4-mm hex screw.
b. Ensure that all tubes are securely connected.
c. Remove the toweling.
d. Ensure that each nozzle is centered above the corresponding cuvette.
7. At the Operation Panel, select Initialize and verify the the DWUD and WUD are in
the up position and that the analyzer is in READY mode.
8. Verify the wash head nozzles are correctly centered in the cuvettes:
a. At the Menu Panel, select
For additional information concerning the Manual Operation window, refer to
the Manual Operation window in the online Operator's Guide.
b. At he Manual Operation window, double-select
c. Select
correctly positioned.
If not, call your local technical support provider or distributor.
d. Verify the nozzles are correctly centered.
e. At the DWUD or WUD window, select
nozzles.
9. At the Operation Panel, select
the up position and the instrument is in the READY state.
110
Maint
to slightly lower the washer nozzles, then verify that they are
Move
Initialize
, then select
Manual Operation
14.DWUD
., then select
Init
, then verify the DWUD and the WUD are in
ADVIA 2400 Operator's Guide
.
or
.
23.WUD
to raise the washer
Exit