Item
Vacuum pump dimensions
Computer dimensions
Monitor dimensions
Optional universal rack handler
dimensions
Weight
Environment
Electrical requirements
Cooling/ventilation requirements
Water requirements
Drain requirements
Appendix
Description
546(h) x 340(w) x 500(d) mm
(21.5 x 13.4 x 19.7 in.)
366(h) x 221(w) x 421(d) mm
(14.4 x 8.7 x 16.6 in.)
422(h) x 417(w) x 447(d) mm
(16.6 x 16.4 x 17.6 in.)
1126(h) x 730(w) x 1022(d) mm
(44.33 x 28.74 x 40.24 in.)
80.7 kg (178 lb)
The ADVIA 2400 Chemistry system is for indoor use at an altitude of
up to 2000 meters with a pollution degree of 2.
Space around the system must be sufficient to allow for ventilation.
There must be a minimum of 19.7 inches (500 mm) behind the system,
32 inches (800 mm) on each side, and 42 inches (1000 mm) in front of
the system.
A 3-kVA power source, single-phase, 2-pole, 3-wire configuration
with Class III grounding.
Can be tapped the following input voltages, (100 V, 115 V, 200 V, 220
V, or 230 V).
Maximum current draw at in-rush is 26 amps at 115 V or 14 amps at
220 V.
Ventilation sufficient to maintain +18°C to +30°C operating
temperature.
The maximum temperature change the system can accommodate is
2°C/hour.
Maximum relative humidity allowable with system operating is 40% to
70% with no condensation.
Heat output:
Power off mode:
0.287 kW
979.3 Btu/hr
246.8 kCal/hr
Ready mode:
1.110 kW
3787.0 Btu/hr
954.4 kcal/hr
Processing mode:
1.635 kW
5579.0 Btu/hr
1406.0 kCal/hr
Deionized (or demineralized) water from a nonpressurized water reservoir
with a 30 liter/hour capability at a pressure of 1.4 to 14 PSI (1 to 10 m of
water). If water pressure is higher, use a pressure regulator.
Minimum of 10.6 gallons (40 liters) per hour.