Replacing The Reaction And Dilution Cuvettes - Siemens ADVIA 2400 Operator's Manual

Chemistry system
Hide thumbs Also See for ADVIA 2400:
Table of Contents

Advertisement

CAUTION
To assist with removal as well as to avoid crimping these thin filter hoses, remove
only 3 or 4 hoses from the water bottle at a time.
6. First unfasten the filter holder from the end of each tube, and then remove the filter.
7. To clean the 10R filter:
a. Place the filter in a beaker filled with a freshly made 10% solution of household
b. After 30 minutes, remove the filter and rinse it with deionized water and replace
CAUTION
Ensure that the filters are properly positioned with the filter holder to avoid filter
shift.
8. Using a lint-free towel soaked in 5% bleach solution, clean the outside surfaces of the
filter holder and hose.
9. Insert the silicon filter hose into the water bottle.
10. Insert the ten Teflon filter hoses in the tank, 3 or 4 at a time.
11. Prime the lines:
a. At the Operation Panel, select the
b. In the PRIME Set dialog box, select
c. Select

Replacing the reaction and dilution cuvettes

Materials required:
• 13 sets of reaction cuvettes (sample
cell RRV, single cuvette set,
REF 05024992, PN 073-0023-02)
• 6 sets of dilution cuvettes (sample cell
DTT, single cuvette set,
REF 05049669, PN 073-0022-01)
Time: 20 minutes
Analyzer mode: OFF
Replace the 20 sets of reaction (RRV) cuvettes and 6 sets of dilution (DTT) cuvettes once
every four months.
140
bleach and water.
onto the filters.
number of times in all fields.
.
Execute
button.
Prime
and type
PRIME 2
10
BIOHAZARD
Wear personal protective
equipment.
Use universal precautions.
ADVIA 2400 Operator's Guide