Preventive Cleaning Of The Wash Station Lines - Siemens ADVIA 2400 Operator's Manual

Chemistry system
Hide thumbs Also See for ADVIA 2400:
Table of Contents

Advertisement

CAUTION
Make a note of the bottle position on the shelf, to avoid mixing up the fluid bottles.
3. Remove the RRV bath oil bottle (
CAUTION
Ensure that level-sensor connector (
4. Refill the bottle with RRV (reaction) bath oil.
5. Replace the bottle on the shelf in the cabinet.
6. Connect the RRV bath oil bottle level sensor connector (
in and turning it clockwise.
7. Insert the filter and tube, then fasten the cap.
NOTE:

Preventive cleaning of the wash station lines

Materials required:
• 5 reagent containers, empty, 70-mL
(optional)
• Probe Wash 1
• Probe Wash 2
• 10% Cuvette Wash solution
• 5% Probe Wash 3 solution
• 70-mL reagent container
(REF 06397121, PN 073-0373-02)
• Wash solution labels (REF 00153468,
PN 073-0406-02)
• Deionized water
Time: 45 minutes
Analyzer mode: READY
If you experience a problem with clogs in the wash station aspiration nozzles and lines,
use this procedure to clean the WUD and DWUD wash station aspiration nozzles and
lines.
Maintenance
Make sure that the filter holder is located at the bottom of the bottle.
).
1
) does not get wet, to avoid damaging it.
2
2
BIOHAZARD
Wear personal protective
equipment.
Use universal precautions.
) by pushing the connector
151