Setting Flow Rates; Checking The Standard Solution 2 (High) Flow Rate - ABB Navigator 550 Operating Instructions Manual

Hydrazine analyzer, measurement made easy
Hide thumbs Also See for Navigator 550:
Table of Contents

Advertisement

Navigator 500
Hydrazine analyzer
Blue
C
Hydrazine
sensor red
connector
(upper)
Hydrazine
temperature
sensor blue
connector
Mixing / delay coil
Fig. 4.1 Setting up the sample flow
20
Red
E
D

4.3 Setting flow rates

Warning. The sample is dosed with sodium hydroxide and
the concentration, although low at first, increases if any
spillage is left to evaporate. Take care to dispose of the
outflow safely.
The flow rates of standard solution and sample through the
wet-section are preset at the factory. It is advisable to check the
flow rates prior to the initial calibration – see Table. 4.1, page 22.
The overflow from the sensor drains into a drain tundish that is
drained to waste via a drain tube.

4.3.1 Checking the standard solution 2 (high) flow rate

The flow rate is factory-set. To check the standard solution
flowrate:
1. Fill the standard solution container with standard solution
or high purity water.
At the transmitter:
2. Access the Calibration level menus by pressing the
key (below the
icon).
Navigator
Tag 1
4.28
4.28
21.7
21.7
N2H4
The Operator menus are displayed:
Navigator
Operator Pages
4.28
4.28
Data Views
Logs
Alarm Acknowledge
Manual Hold
21.7
21.7
Autoscroll
Media Card
Enter Configuration
10 : 34: 52
2012–04–15
ppb
°C
CAL
10 : 34: 52
2012–04–15
ppb
°C
OI/AHM550–EN Rev. A

Advertisement

Table of Contents
loading

This manual is also suitable for:

Navigator 500 seriesNavigator 540

Table of Contents