ABB Navigator 550 Operating Instructions Manual page 23

Hydrazine analyzer, measurement made easy
Hide thumbs Also See for Navigator 550:
Table of Contents

Advertisement

Navigator 500
Hydrazine analyzer
3. Press the
key to select the Enter Configuration menu
and press the
key (below the
The Access Level page is displayed:
Access Level
Read Only
Calibrate
Advanced
Service
Back
4. Use the
key to scroll to the Advanced access level
and press the
key (below the Select prompt) to enter
the top level Configuration menus.
Use the
/
keys to scroll to the Sensor Setup menu
and press the
key (below the Select prompt to enter
the Sensor Setup level menus.
Menu
Sensor Setup
Exit
The Hydrazine 1 (4) page is displayed:
Calibrate
Hydrazine 1
Back
5. Press the
key (below the Select prompt).
The Hydrazine 1 (4) menu page is displayed:
Hydrazine 1
Tag
Units
Range High
Range Low
Filter Type
Filter Time
Solution Detection
Flow Sensor
Man. Valve Control
Back
OI/AHM550–EN Rev. A
icon).
Select
Select
Select
Select
6. Press the
key to select the Man. Valve Control menu
and press the
key (below the Select prompt).
The Man. Valve Control page is displayed with the valve in
the closed position:
Man. Valve Control
Cancel
7. Press the
key to energize (open) the solenoid valve to
enable the flow of standard solution.
Man. Valve Control
Cancel
Note. When the page is exited (by pressing the
key below the Exit prompt), the valve reverts the
previous state (closed).
4.76
ppb
4.76
ppb