Appendix B – Multiple wet-section setup
A single Navigator 500 transmitter can monitor up to 4 wet-sections. The wet-sections can be any combination of the three Navigator
500 parameters – sodium, low level dissolved oxygen and hydrazine. Note that the transmitter cannot monitor more than 1
wet-section if the wet-section is a multi-stream sodium.
If an additional wet-section is added to a transmitter, the procedures in sections B.1 and B.2 (below) must be performed.
B.1 Configuring the device address
The unique device address assigned to the wet-section (1 to 4), enables the transmitter to identify the wet-section on the data
transmission link. Each wet-section must have its own unique address. The address can be set by SW1 as shown in Table B.1 and
Fig. B.1. LEDs D4 and D5 indicate the slave address of the PCB.
Slave address
1
2
3
4
Table B.1 Configuring the device address
SW1.1
OFF position
Configuration DIP switch SW1
Fig. B.1 Configuration DIP switch SW1 and LED D4 / D5 location
OI/AHM550–EN Rev. A
SW1.1
OFF
OFF
ON
ON
Terminals G and W
SW1.2
ON position
ON
1 2
SW1
LED D4 and LED D5
SW1.2
LED D4
OFF
ON
OFF
ON
B R
1 3 5 7 9 1 1 13 15 17 19 21 23 25 27 29 31 33 3
Scr
G W
2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36
D5
D4
LED D5
OFF
OFF
OFF
ON
ON
OFF
ON
ON
67