9 Specification – analyzer
Measuring range
0 to 1000 ppb
Units of measure
ppb, µg/l, µg/kg
Accuracy
±3 % of reading or ±1 ppb, whichever is the greater
Repeatability
±3 % of reading or ±1 ppb, whichever is the greater
Response time
<2 minutes for a 90 % step change
Resolution
0.1 ppb
Temperature compensation
5 to 55 °C (41 to 131 °F) automatic using a Pt1000
AutoCal frequency
Programmable on a daily basis from 1 to 7 days or a weekly basis of
1 to 8 weeks
Sample temperature
5 to 55 °C (41 to 131 °F)
Sample pressure
1.5 bar gauge (21.7 psi) maximum
Sample flow rate
100 to 400 ml/min
Sample connections
1
/
flexible tubing to barbed connector
4
60
Environmental data
Ambient operating temperature:
0 to 55 ºC (32 to 131 ºF)
Ambient operating humidity:
Up to 95 % RH non-condensing
Storage temperature:
– 20 to 70 °C (–4 to 158 °F) without sensor
0 to + 55 °C (41 to 131 °F) with sensor
Approvals, certification and safety
CE mark
UL (pending)
CSA (pending)
General safety
EN61010-1
Pollution category 2
Insulation category 2
EMC
Emissions & immunity
Meets requirements of IEC61326 for an industrial environment and
domestic emissions
Maintenance
Periodic calibration:
User-defined
DS/AHM550–EN
OI/AHM550–EN Rev. A