Navigator 500
Hydrazine analyzer
3.2 Installing the transmitter
3.2.1 Transmitter optional accessories
Optional accessories comprise:
Cable gland kit
3.2.2 Transmitter location
For transmitter general location requirements refer to Fig. 3.6.
Install in a clean, dry, well ventilated and vibration-free location
giving easy access. Avoid rooms containing corrosive gases or
vapors, for example, chlorination equipment or chlorine gas
cylinders.
Warning. The transmitter is not fitted with a switch – an
isolation device such as a switch or circuit breaker
conforming to local safety standards must be fitted to the
final installation. It must be fitted in close proximity to the
transmitter, within easy reach of the operator and marked
clearly as the isolation device for the transmitter.
Eye-level location
Ambient temperature
Fig. 3.6 Transmitter location
12
55 °C
(131 °F
Max.
– 10 °C
(14 °F)
Min.
Humidity
0 to 95 %
Avoid vibration
3.2.3 Panel mounting
Dimensions in mm (in.)
Mounting panel –
maximum thickness
6 (0.236)
40 (1.57)**
Panel cut-out dimensions*
* To DIN43700
**150 mm (6 in.) if (optional) cable glands fitted
Fig. 3.7 Transmitter panel-mount option
92
35
(1.4)
(3.6)
Cable gland kit
– optional
Panel-mount kit –
optional
+1.1
186
–0
(7.32
+0.04
)
–0
+1.1
186
–0
(7.32
+0.04
–0
40 (1.57)
OI/AHM550–EN Rev. A
)