Installation; Installing The Wet-Section; Sample Requirements; Location - ABB Navigator 550 Operating Instructions Manual

Hydrazine analyzer, measurement made easy
Hide thumbs Also See for Navigator 550:
Table of Contents

Advertisement

Navigator 500
Hydrazine analyzer

3 Installation

3.1 Installing the wet-section

3.1.1 Sample requirements

Ensure the sampling point is as close as possible to the
wet-section and provides a thoroughly-mixed representative
sample.
Sample must contain less than 10 ppm suspended solids
with a particle size no greater than 60 µm. (If particle sizes
exceed 60 µm, use a 60 µm filter.)
Sample temperature must be within the range 5 to 55 °C
(41 to 131 °F).
Sample flow rates must be within the range 100 to
400 ml/min (6.01 to 24.40 cu in./min).
Sample must be at atmospheric pressure.
8

3.1.2 Location

For general location requirements refer to Fig. 3.1. Install in a
clean, dry, well ventilated and vibration-free location giving easy
access. Avoid rooms containing corrosive gases or vapors, for
example, chlorination equipment or chlorine gas cylinders.
Select a location away from strong electrical and magnetic
fields. If this is not possible, particularly in applications where
mobile communications equipment is expected to be used,
screened cables within flexible, earthed metal conduit must be
used.
Maximum cable distance
transmitter to
wet-section 30 m (98 ft.)
0 °C
(32 °F)
Min.
Navigator 500
Avoid vibration
Fig. 3.1 Locating the wet-section
Navigator 500
55 °C
(131 °F)
Max.
Humidity
IP54
0 to 95 %
OI/AHM550–EN Rev. A