Replacing the Top Cover Assembly

Lift the front edge of the top cover and disconnect the spade connector for the ESD ground wire from
4
the terminal on the metal chassis.
5
Now raise the cover a little more, and disconnect the multipin connector of the ribbon cable from the
MIF board. Pull straight up to avoid bending the pins. The top cover assembly is now free of the chassis
assembly.
Replacing the Top Cover Assembly
The serial number appears initially in two places on the base station - on the rear of the top cover
assembly, and on the bottom of the metal chassis. A replacement top cover does not carry a serial
number.
Ensure all items are replaced in the top cover assembly.
1
Reconnect the multipin ribbon cable connector to the socket on the MIF board.
2
Ribbon cable
connector
4 Disassembly/Reassembly
ESD ground
spade connector
Top
cover
assembly
Chassis
assembly
Reconnect ribbon cable connector
29