4 Disassembly/Reassembly
Step 6: Replacing the MIF Board
Carefully locate the multipin connector plug, on the underside of the MIF board, in the socket on the
1
receiver board, ensuring it is accurately aligned.
2
Carefully apply a little pressure on the left hand rear corner of the MIF board to seat the multipin
connector into the socket on the receiver board below.
Ensure that the screw holes in the board align with the screw fixing points below.
3
Drive in the two screws (B) with a T-10 Torx driver.
4
5
Then drive in the two screws (A) using a T-10 Torx driver. The MIF board is now secure.
38
Changing the Power Supply/Metal Chassis Assembly
Screws (B)
Screws (A)
Multipin connector