S(3): Enclosure Leakage Current Test - Single Fault Condition (Sfc) Open Supply; S(4): Enclosure Leakage Current - Sfc Open Earth (Ground) - Philips Avalon CTS M2720A Service Manual

Cordless fetal transducer system
Hide thumbs Also See for Avalon CTS M2720A:
Table of Contents

Advertisement

Safety Tests
S(3): Enclosure Leakage Current Test
- Single Fault Condition (SFC) Open Supply
Test to perform:
The enclosure leakage current: single fault condition open supply is applicable to Class 1 and 2 equipment,
type B, BF, and CF Applied Parts. The test measures leakage current of exposed metal parts of Instrument
Under Test with one supply lead interrupted; it tests normal and reversed polarity.
For type BF and CF Applied Parts measures AP/GND.
Safety Test according IEC 60601-1 (Clause 19.4g).
Report the highest value.
S(4): Enclosure Leakage Current
- SFC Open Earth (Ground)
Test to perform:
The enclosure leakage current: single fault condition open earth (ground) is applicable to Class 1
equipment, type B, BF and CF Applied Parts. The test measures leakage current of exposed metal parts of
Instrument Under Test with Protective Earth open-circuit; it tests normal and reversed polarity.
For type BF and CF Applied Parts the test measures AP/GND.
Safety Test according IEC 60601-1 (Clause 19.4g).
Report the highest value.
L (N)
S2
S1
N (L)
S4
(*)
PE
MD
6 Preventive Maintenance and Safety
Medical electrical system
Instrument under test
Signal parts
in- and/or
output
Applied part
(*) Not present in Class 2.
(**) Can be multiple different connections to different equipment at same time.
Signal parts
(**)
in- and/or
output
Other
Instrument
53