Changing the Power Supply/Metal Chassis Assembly

Reconnect the power supply cable connector to the power supply.
8
Replace the top cover assembly (see page 29).
9
Changing the Power Supply/Metal Chassis Assembly
The power supply and the metal chassis come as one unit (they are not available independently). The
power supply cable is reusable, and also comes with the new chassis/power supply assembly.
To change the power supply/metal chassis assembly:
Step1: Remove the Top Cover Assembly
See page 28.
Step2: Remove the MIF Board
Remove the six-pin power supply connector from the MIF board.
1
Six-pin power
connector
2
Remove the two screws (A) using a T-10 Torx driver.
Screws (A)
4 Disassembly/Reassembly
33