Network Test

Auxiliary Port Test

Remote Control Test
Revision A
Maintenance: Checkout Procedures
127(
Some Dash monitors do not support the docking station's Ethernet or
auxiliary communication links. See
page 2-9
and
"Compatible Auxiliary Devices" on page
127(
To complete this test, there must be at least one other bed located on
the network.
1. Verify that a monitor is connected to the docking station.
2. Verify the docking station's communication indicator is illuminated
green.
3. Verify the docking station's security lever is in the far left position.
4. Verify that the docking station (not the monitor) is connected to the
Unity-MC (Mission Critical) network.
5. Press the Power key on the monitor to turn it on and wait for all of
the parameter boxes to display on the monitor.
6. Select MORE MENUS > VIEW OTHER PATIENTS > SELECT
ANOTHER CARE UNIT.
7. Verify that you can see at least one care unit.
8. Select a care unit.
9. Select SELECT A BED TO VIEW.
10. Select a bed.
11. Verify that the patient window appears on the monitor's split-screen.
Complete the remote control test to verify the docking station's two
auxiliary connectors are functioning.
1. Verify the remote control is connected into an Autoport-to-Mport
adapter and is inserted into the docking station's Aux connector.
2. From the monitor, select MORE MENUS -> MONITOR SETUP ->
REVISION AND ID.
3. Select NEXT from the popup menu to display the port connectors.
4. Verify the REMOTE CONTROL label appears after the appropriate
port and the software version for the remote control is 1A.
5. Press each remote control key and verify a beep tone sounds at the
monitor.