GE Dash Port 2 Service Manual page 71

Table of Contents

Advertisement

Problem
The monitor pivots when the security lever
is in the lock position.
The docking station will not release the
monitor when the security lever is moved to
the unlock position.
The security lever doesn't move to the far
left position.
Revision A
Troubleshooting: Troubleshooting Procedures
Problems and Solutions Table
Cause
The docking station's locking mechanism
may have been accidentally activated
before the monitor was placed on the
docking station.
The monitor might not be positioned
correctly on the docking station to allow all
four clamps to secure the monitor's GCX
plate.
The Top Assembly FRU has failed.
The security lever was not moved and held
in the unlock position while the monitor was
removed.
The Top Assembly FRU has failed.
Cables or other objects may be blocking
the monitor's AC power or communication
ports.
The monitor's expansion PCB may be
damaged.
The monitor may not be positioned
correctly on the docking station.
Foreign objects may be blocking the
monitor's expansion PCB.
The Connector Assembly FRU has failed.
Dash Port 2 Docking Station
2000966-138
Solution
1. Remove the monitor.
2. Re-set the locking mechanism by
moving the security lever to the unlock
(far right) position.
3. Reposition the monitor.
1. Remove and re-position the monitor on
the docking station.
2. Verify the monitor doesn't pivot.
"Replace the Top Housing FRU"
page 6-18.
Slide and hold the security lever in the
unlock (far right) position until you remove
the monitor from the docking station.
1. Remove the six screws from the bottom
of the docking station to separate the top
housing of the docking station from the
bottom housing of the docking station.
2. Use a bladed screw driver to manually
release the docking station's clamps
from the base of the monitor.
3. If necessary, contact Technical Support.
4.
"Replace the Top Housing FRU"
page 6-18.
1. Remove the monitor from the docking
station.
2. Remove any cables or other objects
from the connector ports located on the
bottom of the monitor.
3. Reposition the monitor.
Replace the monitor's damaged expansion
PCB. See the Dash Patient Monitor's
Service Manual.
Remove and reposition the monitor.
1. Remove foreign objects blocking the
expansion PCB.
2. Reposition the monitor.
See,
"Replace the Connector Assembly
FRU or PCB FRU"
on page 6-19.
on
on
5-13