Problem
No picture appears on the remote display
and the following conditions apply:
The remote display's illuminated LED
is amber.
The docking station's communication
indicator is either illuminated green
OR is NOT illuminated at all.
The security lever is already in the central
position BEFORE the monitor has been
placed on the docking station.
The security lever doesn't move to the
central position when the monitor has been
placed on the docking station.
5-12
Troubleshooting: Troubleshooting Procedures
Problems and Solutions Table
Cause
The docking station or the monitor is
defective.
The locking mechanism may have been
accidentally activated before the monitor
was placed on the docking station.
The monitor's footpad and/or GCX plate
may have been removed. Without both
components, the monitor WILL NOT safely
or correctly function with the docking
station.
Cables or other objects may be preventing
the monitor from activating the locking
mechanism.
The monitor might not be positioned
correctly on the docking station to activate
the locking mechanism.
The Top Assembly FRU has failed.
Dash Port 2 Docking Station
2000966-138
Solution
1. Remove the monitor from the docking
station.
2. Manually depress the locking lever pin
and slide the security lever to the far left
position.
3. Verify the AC power/communication
mechanism extends out of the docking
station's base and locks into position.
4. If the AC power/communication
mechanism does NOT extend out of the
docking station's base, see
"The AC
power/communication mechanism does
NOT extend out of the docking station's
base."
on page 5-10.
5. If the AC power/communication
mechanism extends out of the docking
station's base, then complete the
following tasks.
a. Replace the docking station with a
known good docking station.
b. If a picture appears on the remote
display, then replace the PCB in the
defective docking station. See
"Replace the Connector Assembly
FRU or PCB FRU"
on page 6-19.
c. If a picture still does NOT appear on
the remote display, replace the
monitor.
1. Remove the monitor.
2. Re-set the docking station's locking
mechanism by moving the security lever
to the unlock (far right) position.
Replace the missing footpad and/or GCX
plate. See the Dash Patient Monitor's
Service Manual.
Verify the docking station's base is free of
objects.
Remove and re-position the monitor.
"Replace the Top Housing FRU"
page 6-18.