[5] Verifying the adjustment results
(1) Have on hand IPs of all sizes that are employed at the user site. Expose these IPs to 1
mR of radiation.
(2) Using "Sensitivity" on the IDT's test menu, record and output images.
(3) Note the generated film outputs to verify that either the top edge or the bottom edge of
each IP is invisible.
GOOD
009-058-00
10.20.2000 FM2732 (2)
Edge
Edge
CR-IR347 Service Manual
Edge
NG
FR7H3741.EPS
MC - 252