I Check procedure
(1) Turn ON the power.
(2) Start the M-Utility.
(3) Perform SK2 operational checkout.
#1 [Input/ENT]
#2 [Input/ENT]
#3 [Input/ENT]
REFERENCE
For SK2 operational checkout, perform the following steps:
1. Insert a cassette all the way into the cassette set unit.
¡ Check that the SK2 closes.
2. Try to pull the cassette while it is locked with the hold pin.
¡ Check that the SK2 remains closed.
3. Remove the cassette.
¡ Check that the SK2 opens.
009-058-00
10.20.2000 FM2732 (2)
0. QUIT
1. ERROR LOG UTILITY
2. CONFIGURATION SETTING
3. TEST MODE
4. ELECTRICAL UTILITY
5. SCANNER UTILITY
6. MECHANICAL UTILITY
7. FILE UTILITY
8. BACKUP MEMORY
9. HV OFF
10. MENU SETTING
11.SYSTEM UTILIT
Y
>
6
0. QUIT
1. MOTOR
2. ACTUATOR
3. SENSOR
4. UNIT
MU >
3
0. QUIT
1. NUMBER
2. MONITOR
3. MONITOR ALL
MU : SEN >
3
CR-IR347 Service Manual
MC - 47
FR7H3330.EPS
MC - 47