(4) Verify the SK2 assembly mounting position.
150 mm scale
009-058-00
10.20.2000 FM2732 (2)
A
SK2
DETAIL B
Actuator
SK2
CR-IR347 Service Manual
B
DETAIL A
Guide plate
[Check]
SK2
SK2
The SK2 should be OFF
when this gap is about 1.0 mm.
SK2
The SK2 should be ON
when this gap is about 1.5 mm.
MC - 48
Scale
Scale
Guide plate
FR7H3379.EPS
MC - 48