I Installation procedure
To accomplish installation, reverse the removal procedure.
I Check/adjustment procedure
• Verify that the timing belt is properly threaded through its path.
• After the timing belt is replaced with a new one or the tensioner is removed and reinstalled,
remove the IP removal unit and adjust the belt tension with the tensioner screw.
"5.1 IP Removal Unit"
361 mm
Timing belt
009-058-00
10.20.2000 FM2732 (2)
Scales
CR-IR347 Service Manual
Push-pull gauge
560.5 N
MC - 88
FRONT
FR7H3208.EPS
MC - 88