I Installation procedure
N NOTES N
• When installing the SYN08A assembly, properly position it by pressing it against the
reference surface of the positioning bracket.
Scanning optics unit
• Use the existing spacers for the newly installed assembly.
• Position the cable in the cable guide.
Accomplish installation by reversing the removal procedure.
I Check procedure
(1) Install all covers.
(2) Turn ON the machine.
Verify that the machine starts up normally.
(3) Adjust the format.
"9.1 Adjusting the Format"
(4) Check for image problems.
"9.3 Checking for Image Problems"
009-058-00
10.20.2000 FM2732 (2)
DETAIL A
SYN08A assembly
CR-IR347 Service Manual
A
Mounting bracket
Reference surface
FR7H3717.EPS
MC - 276
MC - 276