Using the Pacer
15 To set the lowest possible output level to capture, decrease the current by decre-
ments of 5 mA by pressing Output.
NOTE
If the monitoring ECG lead falls off while pacing in demand mode, the pacer will
stop delivering pulses and the messages PACER STOP and LEADS OFF will
appear. To resume pacing, reattach the lead and press Start/Stop.
Pacing in fixed mode does not require ECG leads to be attached for the pacer to
deliver pulses.
If a multifunction defib electrode comes off during pacing, the pacer will stop
delivering pulses and the messages PACER STOP and PADS OFF will appear. To
resume pacing, reattach the electrode and press Start/Stop.