Charging The Battery - HP Codemaster 100 User Manual

Portable defibrillator/monitor
Table of Contents

Advertisement

CodeMaster 100

Charging the Battery

Insert battery into the HP M2480B battery support system charging bay. The
The battery is fully charged when the
discharged HP M2476B 2.5 amp-hour battery takes approximately 2 hours to
charge (a fully discharged HP M2477B 4.0 amp-hour battery takes
approximately 3 hours to charge).
It may take longer to charge the battery pack if the temperature of the battery
is not between 10°C and 40°C. If it is outside these temperatures, the battery
support system allows the battery to warm up or cool down before charging.
A flashing
to the indicator. Notify service personnel and do not use that charging bay.
The other charging bay is not affected and you can continue to use it.
If the
WARNING
battery from service immediately and notify service personnel.
If the
NOTE
recommended.
13-14
CHARGING
indicator appears, to indicate that the battery is charging.
FAULT
indicator indicates a problem with the charging bay next
indicator appears, the battery did not charge. Remove the
FAULT
START
indicator (
READY
indicator appears. A fully
) flashes, a battery pack capacity test is