If The Defibrillator Does Not Charge - HP Codemaster 100 User Manual

Portable defibrillator/monitor
Table of Contents

Advertisement

Table 12-2
Defibrillator Messages (Continued)
Message
CHECK
BATTERY
CHECK
SETUP

If the Defibrillator does not Charge

If the defibrillator does not charge, perform the following steps:
1 Verify the proper setting of the Energy Select control.
2 If it is correct, follow these steps.
3 If the instrument remains unable to charge, turn the Energy Select con-
4 Press Review to print an event summary and keep any ECG strips from
5 Alert HP service personnel.
Cause
a.
Turn the Energy Select control to Off , and then back to the desired
energy setting.
b. Press Charge again.
trol to Off and use a backup defibrillator.
the defibrillator for later evaluation.
CodeMaster 100
Possible Solutions