Defibrillator/Monitor
Thermal Array Recorder
Event Summary: Stores and prints 3 seconds pre- and 8 seconds post-critical event
data for up to 28 events. Data retained after unit is turned off.
Annotates: Time, date, HR, SpO
, event marker, ECG mode, defibrillator mode,
2
selected energy, actual delivered energy, peak current, patient impedance, and
pacing current and rate.
Speed: 25 mm/sec.
Paper Size: 50 mm by 30 m (100 ft).
Recorder Mode: Automatically documents events and ECG during defibrillation
episodes. The recorder can be configured to run in either real time or with a six
second delay.
Frequency Response:
-CodeMaster 100 without the 12-Lead option: 0.5 to 40 Hz
-CodeMaster 100 with the 12-Lead option: 0.05 to 150 Hz selectable.
Size and Weight of the CodeMaster 100 without the 12-Lead Option
Defibrillator dimensions: 6.25 in H by 13.75 in W by 15.125 in D (15.9 cm H by
34.9 cm W by 38.4 cm D).
Defibrillator weight: 21.5 lb (9.75 kg). Includes external paddles, battery, and
recorder paper.
14-5