The HP CodeMaster 100 stores high voltage energy and is capable of
delivering up to 360 joules of DC energy to a 50 ohm impedance.
To remove power from the instrument, you must turn the Energy Select
control to Off.

Disarming the Defibrillator

There are three ways to disarm a charged instrument:
• Turning the Energy Select control from an energy level setting to the
• Placing the paddles in their holders and depressing both shock (
• Waiting for the instrument to automatically disarm. The unit disarms auto-
Do not leave the instrument turned on when it is not in use.
CAUTION
Do not discharge the defibrillator with the paddles shorted together which
can cause burning and pitting of the metal paddle contacts.
Disconnect any other medical electronic equipment from the patient during
defibrillation unless labeled as defibrillator protected (

Safety Considerations

Monitor On position.
buttons simultaneously.
matically after it has been left charged for 60 seconds in the Manual mode
and after 30 seconds in the Advisory Mode.
Safety Considerations
)
) (
).
1-21