Start-up
SERVICE
This menu item is protected with a password. Access is only allowed to BEKO SERVICE or to authorised skilled
personnel.
SCIENTIFIC MODE (SPECIAL MODE)
When this function is activated, measurement takes place via calibration gas connection "C" and zero air connection
"Z". The measuring gas connection "M" needs to be closed. The "PURIFICATOR" reference gas generation unit is not
used in this mode. In the "Measurement" menu, when selecting the gas type, a warning will inform you that this mode is
activated.
NOTE!
Please use this mode only after prior consultation with BEKO. Close measuring gas connection "M". Connect
zero air to "Z". Connect measuring gas to "C".
The max. permissible operating pressure in the scientific mode is 3 bar(g).
Now confirm your settings with OK.