Operation
Example setting 2
If the residual oil content values indicated in the METPOINT OCV
0.1 mg/m³ and ≤ 1.0 mg/m³, please enter "0.01" into the blue field and confirm with OK.
Scaling
Scaling initial value
Scaling final value (Full-Scale)
Example setting 3
If the residual oil content values indicated in the METPOINT OCV
1.0 mg/m³ and ≤ 5.0 mg/m³, please enter "0.1" into the blue field and confirm with OK.
Scaling
Scaling initial value
Scaling final value (Full-Scale)
For all other measured-value ranges, the 4-20 mA signal can be scaled using the same steps.
64
0.1 mA
corresponds to
4.0 mA
corresponds to
20 mA
corresponds to
0.1 mA
corresponds to
4.0 mA
corresponds to
20 mA
corresponds to
®
MEASUREMENT mode lie in the range between ≥
0.01
mg/m³
0.00
mg/m³
1.60
mg/m³
®
MEASUREMENT mode lie in the range between ≥
0.1
mg/m³
0.0
mg/m³
16
mg/m³
METPOINT
®
OCV