4.9.3 Viscometer assembly procedure

(A) Replace the cylinder, and then insert the plunger. Handle the plunger with tweezers
making sure that no dust adheres to the surface of plunger.
(B) Set the collar to cylinder, taking care with its orientation (with the convex up).
(C) Make sure that an O-ring is set in the viscometer cover, and then assemble the cover
into the viscometer case.
(D) Attach the three viscometer setscrews and tighten them. (Tighten the screws with
reasonable force using screwdriver. Tightening the screws excessively will break the
threads.)
O-ring
Plunger
Cylinder
Coil

4.9.4 Viscometer attaching procedure

(A) Set the O-ring and squeezing pin on the viscometer OUT side, and use the two
setscrews to attach them to the circulation unit.
(B) Plug in the connectors of lead wires again.
(C) Connect the viscometer IN-side tube (mark G).
Note: After reassembly, it may be necessary to replace the ink and/or calibrate
the viscometer.
When disassembling and washing during installation:
When replacing the viscometer during installation:
When replacing the viscometer in cases other than installation:
Viscometer case
Viscometer cover
Viscometer IN joint
Collar
Calibrate the viscometer without replacing the ink.
Calibrate the viscometer without replacing the ink.
First replace the ink, then calibrate the viscometer.
4-21
Fixing
plate
Thermistor
O-ring
O-ring
O-ring
Squeezing pin