Step
5
6
ÄKTA avant User Manual 29035184 AE
Action
In the Process Picture:
Click the System pumps icon.
Enter a low System flow (e.g., 0.2 ml/min).
Click Set flow rate.
Result: A system flow of 0.2 ml/min starts.
When buffer leaves the tubing on port 1A (if port 1A was chosen in the
method to be run) in a continuous mode and the top part of the column is
filled with buffer, connect the tubing to the top of the column.
5 Operation
5.5 Connect a column
185