8.6
Troubleshooting: Pumps
In this section
•
•
•
Pumps
Problem
Liquid is leaking be-
tween the pump head
and the side panel
Low eluent flow and
noise
ÄKTA avant User Manual 29035184 AE
Troubleshooting for System pumps and Sample pumps
Example of pump pressure curves
Remove persistent air bubbles
Possible cause and action
Piston seal or rinsing membrane incorrectly fitted
•
or worn
Replace or reinstall the seal or the membrane. See
ÄKTA avant Operating Instructions.
Air in pumps
•
Purge the pumps. See Section 5.4.1 Prime buffer inlets
and purge system pumps, on page 166.
Bad piston spring
•
Disassemble the pump head and examine the piston
spring. If the spring is corroded, check piston seal and
rinse membrane. Make sure that the pump piston
rinsing system is always used when working with
aqueous buffers containing salt.
Bad pump piston
•
If the piston is damaged, replace it according to ÄKTA
avant Operating Instructions.
Bad pump piston seal
•
Replace the piston seal and rinse membrane accord-
ing to ÄKTA avant Operating Instructions.
Blocked pump flow restrictors
•
See Section 7.6.10 Clean pump flow restrictors, on
page 315.
8 Troubleshooting
8.6 Troubleshooting: Pumps
397