HP 81101A Reference Manual page 49

50 mhz
Hide thumbs Also See for 81101A:
Table of Contents

Advertisement

Value coupling
Range coupling
Offset
Description
Use this command to program the amplitude current of the OUTPUT
signal. Note that to set the OUTPUT levels in terms of current, you first
have to execute the [:SOURce]:HOLD CURRent command to enable the
[:SOUR
The available current range is limited by the combination of:
• Specified Voltage limits
• Actual OUTPUT Impedance setting :OUTPut:IMPedance
• Actual Expected Load impedance setting:
Example
To program the amplitude current of the output signal:
:HOLD CURR
:CURR 75MA
Command
:CURR[1]:OFFSet
Long
[:SOURce]:CURRent[1][:LEVel][:IMMediate]:OFFSet
Form
Set & Query
Parameter
Numeric
Parameter suffix
A with engineering prefixes.
0.0 µA (50 Ω into 50 Ω)
*RST value
Value coupling
Amplitude = High – Low
High – Low
Offset =
]:CURR
subsystem.
CE
ENT
:OUTPut:IMPedance:EXTernal
Amplitude = High – Low
High – Low
Offset =
2
Enable CURRENT subsystem
Set OUTPUT amplitude to 75 mA
2
Programming Reference
SCPI Instrument Command List