HP 81101A Reference Manual page 84

50 mhz
Hide thumbs Also See for 81101A:
Table of Contents

Advertisement

Programming Reference
SCPI Instrument Command List
Command
:SYST:WARN:STR?
Long
:SYSTem:WARNing:STRing?
Form
Query
*RST value
Not applicable
Description
Use this command to read all the currently active warning messages. The
warning messages are concatenated to form a single string with a ";" as
separator between the messages.
Command
:SYST:WARN:BUFF?
Long
:SYSTem:WARNing:BUFFer?
Form
Query
*RST value
Not applicable
Description
Use this command to read the maximum possible number of characters
that could be returned by :SYST:WARN:STR? if all warnings were active.
Command
:TRIG:COUNt
Long
:TRIGger[:SEQuence[1]]:COUNt
Form
Set & Query
Parameter
Numeric
*RST value
1
Specified limits
1 to 65 536
Description
Use this command to set/read the number of trigger events (pulse
periods) to be generated for each arming event. This corresponds to
selecting the event mode on the M
PULSES
BURST of
84
ODE
Set a trigger count to 1 so that a single pulse period is generated for
each arming event. The instrument is in pulse (stream) mode.
Set a trigger count of 2 to 65536 so that a burst of 2 to 65536 pulse
periods is generated for each arming event.
/T
:
RIGGER SCREEN