HP 81101A Reference Manual page 74

50 mhz
Hide thumbs Also See for 81101A:
Table of Contents

Advertisement

Programming Reference
SCPI Instrument Command List
Command
:VOLT[1]:LIM
Long
[:SOURce]:VOLTage[1]:LIMit[:HIGH]
Form
Set & Query
Parameter
Numeric
Parameter suffix
V with engineering prefixes.
*RST value
+500 mV
Description
Use this command to set/read the high level voltage limit. If you switch
on voltage limiting, the high level voltage cannot be set above the
programmed limit. Note that the voltage is not limited by the OUTPUT
hardware, this is a software limit.
Example
To set and activate the high level voltage limit:
:HOLD VOLT
:VOLT:LIM 3V
:VOLT:LIM:STAT ON
Command
:VOLT[1]:LIM:LOW
Long
[:SOURce]:VOLTage[1]:LIMit:LOW
Form
Set & Query
Parameter
Numeric
Parameter suffix
V with engineering prefixes.
*RST value
–500 mV
Description
Use this command to set/read the low level voltage limit. If you switch on
voltage limiting, the low level voltage cannot be set below the
programmed limit. Note that the voltage is not limited by the OUTPUT
hardware, this is a software limit.
Example
To set and activate the low level limit:
:HOLD VOLT
:VOLT:LIM:LOW 0V
:VOLT:LIM:STAT ON
74
Enable VOLTAGE subsystem
Set OUTPUT high level limit to 3 V
Switch on OUTPUT limits
Enable VOLTAGE subsystem
Set OUTPUT low level voltage
Switch on OUTPUT limits

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents