HP 81101A Reference Manual page 94

50 mhz
Hide thumbs Also See for 81101A:
Table of Contents

Advertisement

Specifications
HP 81101A Specifications
Width
Can be entered as absolute width, duty cycle or trailing edge delay.
Width
Range:
Accuracy:
Duty cycle:
Duty cycle values from 0.1% to 95% can be entered directly. For values
>95 % press SHIFT and use the Modify knob. Note that pulses may be
deteriorated or skipped due to the inaccuracy of period and width. Hence
for large values, it is better to select complement and enter 100 minus the
required duty cycle value.
Delay
Measured between trigger output and main output. Can be entered as
absolute delay, phase ° or % of period.
Delay
Fixed delay from
TRIGGER OUT:
Additional variable range:
Accuracy:
94
HP 81101A
10 ns to 999.5 s
(max value: period –10 ns)
±5 % ±250 ps
0.1% to 95% (depends on period and width;
overprogrammable to 99%)
HP 81101A
17.0 ns typical
0.00 ns to 999.5 s
(max value: period –20 ns)
±5 % ±1 ns

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents