HP 81101A Reference Manual page 51

50 mhz
Hide thumbs Also See for 81101A:
Table of Contents

Advertisement

have to execute [:SOURCE]:HOLD CURRent command to enable the
[:SOURCE]:CURRent subsystem.
The available current range is limited by the combination of:
• Specified Voltage limits
• Actual OUTPUT Impedance setting :OUTPut:IMPedance
• Actual Expected Load impedance setting:
Example
To program the high level current of the output signal:
:HOLD CURR
:CURR:HIGH 150MA
Command
:CURR[1]:LOW
Long
[:SOURce]:CURRent[1][:LEVel][:IMMediate]:LOW
Form
Set & Query
Parameter
Numeric
Parameter suffix
A with engineering prefixes.
Value coupling
Range coupling
High level
–10 mA (50 Ω into 50 Ω)
*RST value
Specified Limits
10 V Outputs (from high Z into short): –400 mA to 396 mA typical
3.8 V Outputs (from 50 Ω into short): –84 mA to 150 mA typical
Description
Use this command to program the low level current of the OUTPUT
signal. Note that to set the OUTPUT levels in terms of current, you first
have to execute the [:SOURce]:HOLD CURRent command to enable
the [:SOURce]:CURRent subsystem.
The available current range is limited by the combination of:
:OUTPut:IMPedance:EXTernal
Amplitude = High – Low
High – Low
Offset =
Enable CURRENT subsystem
Set OUTPUT high level to 150 mA
2
Programming Reference
SCPI Instrument Command List