HP 81101A Reference Manual page 57

50 mhz
Hide thumbs Also See for 81101A:
Table of Contents

Advertisement

Command
:PULS:DCYC[1]
Long
[:SOURce]:PULSe:DCYCle[1]
Form
Set & Query
Parameter
Numeric
Value coupling
Width =
*RST value
10.0% (derived from width and period)
Specified limits
0.001% to 99.9%, depends on width, transition & period.
Description
Use this command to program the duty cycle of the pulse signal. If you
want to set an absolute pulse width use [:SOURce]:PULSe:WIDTh[1].
If you want the pulse duty cycle to remain constant when the pulse
period is varied (rather than the absolute pulse width use)
:SOURce]:PULSe:HOLD[1] DCYCle
Example
To set and hold the duty cycle:
:PULS:DCYC 25PCT
:PULS:HOLD DCYC
Command
:PULS:DEL[1]
Long
[:SOURce]:PULSe:DELay[1]
Form
Set & Query
Parameter
Numeric
Parameter suffix
S with engineering prefixes. You can change the default unit using
[:SOURce]:PULSe:DELay[1]:UNIT.
Value coupling
*RST value
0.0
Duty Cycle
100
Delay
Phase =
Period
Delay
Delay% =
Period
SCPI Instrument Command List
×
Period
Set OUTPUT duty cycle to 25%
Hold duty cycle constant with varying period
× 360
× 100
Programming Reference
57