GE LOGIQ E9 User Manual page 123

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

Basic Measurements (continued)
Table 2-6:
Measurement
Depth
Distance:
Axial (Equal to
1540m/s)
Lateral (Equal to
1540m/s)
Lateral (Equal to
1540m/s)
Lateral (Equal to
1540m/s)
Axial/Lateral (Not
Equal to 1540m/s)
Circumference:
Trace (Equal to
1540m/s)
Ellipse (Equal to
1540m/s)
Trace (Not Equal
to 1540m/s)
Ellipse (Not Equal
to 1540m/s)
Area:
Trace (Equal to
1540m/s)
Ellipse (Equal to
1540m/s)
Trace (Not Equal
to 1540m/s)
Ellipse (Not Equal
to 1540m/s)
Static 3D/4D Volume
Accuracy
Time
LOGIQ E9
User Guide
Direction 5454884-100 English Rev. 1
System Measurements and Accuracies
Units
Useful Range
mm
Full Screen
mm
Full Screen
mm
Full Screen
mm
Full Screen
mm
Full Screen
mm
Full Screen
mm
Full Screen
mm
Full Screen
mm
Full Screen
mm
Full Screen
Full Screen
2
mm
2
Full Screen
mm
2
Full Screen
mm
2
Full Screen
mm
s
Timeline Display
Measurement and Analysis
Accuracy
±20%
±3% or ±1mm,
whichever is greater
±5% or ±1mm,
whichever is greater
±6.5% or ±3.5mm,
whichever is greater
±5% or ±1mm,
whichever is greater
±7.5% or ±5.0mm,
whichever is greater
±5% or ±1mm,
whichever is greater
±5% or ±1mm,
whichever is greater
±7.5% or ±5mm,
whichever is greater
±7.5% or ±5mm,
whichever is greater
±10% or ±5mm
2
,
whichever is greater
±10%, or ±5mm
2
,
whichever is greater
2
±20% or ±20mm
,
whichever is greater
±20% or ±2mm
2
,
whichever is greater
±13%
±5%, not to exceed
10ms
Limitations or
Conditions
Linear Probes
Convex Probes
Sector Probes
All probes Not
Equal To 1540m/s
All probes Not
Equal To 1540m/s
All probes Not
Equal To 1540m/s
All probes Not
Equal To 1540m/s
All probes Not
Equal To 1540m/s
M or Doppler
Mode
2-37

Advertisement

Table of Contents
loading