Declaration of Immunity (continued)
Immunity Type
IEC 61000-4-2
Static discharge (ESD)
IEC 61000-4-4
Electrical fast
transient/burst
IEC 61000-4-5 Surge
Immunity
IEC 61000-4-11
Voltage dips, short
interruptions and
voltage variations on
mains supply
IEC 61000-4-8
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-6
Conducted RF
IEC 61000-4-3
Radiated RF
IEC 61000-3-2
Harmonic Disturbance
Emissions
IEC 61000-3-3 Voltage
Fluctuations Emissions
NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people. If noise generated from other electronic
equipment is near the probe's center frequency, noise may appear on the image. Good power line isolation
is required.
LOGIQ E9
User Guide
Direction 5454884-100 English Rev. 1
Table 4-6:
Declaration of Immunity
Equipment
Capability
± 9 kV contact
± 12 kV air
± 2 kV for mains
± 1 kV for SIP/SOP
± 1 kV differential
± 2 kV common
< 50
(> 95% dip) for
T
0.5 cycle;
400
(60 0ip) for 5
T
cycles;
700
(30 0ip) for 25
T
cycles;
< 50
(>95% dip) for 5
T
sec
6 A/m
3 V
RMS
150 kHz - 80 MHz
3 V/m
80 MHz - 2.5 GHz
IEC Limits
IEC Limits
Regulatory
Acceptable Level
± 9 kV contact
± 12 kV air
± 2 kV for mains
± 1 kV for SIP/SOP
± 1 kV differential
± 2 kV common
< 50
(> 95% dip) for
T
0.5 cycle;
400
(60 0ip) for 5
T
cycles;
700
(30 0ip) for 25
T
cycles;
< 50
(>95% dip) for 5
T
sec
6 A/m
3 V
RMS
150 kHz - 80 MHz
3 V/m
80 MHz - 2.5 GHz
230Vac
230Vac 50Hz, 100Vac
50Hz
Safety Precautions
EMC Environment
and Guidance
Floors should be wood,
concrete, or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Mains power quality
should be that of a typical
commercial and/or
hospital environment. If
the user requires
continued operation
during power mains
interruptions, it is
recommended that the
system be powered from
a UPS or a battery.
NOTE: UT is the a.c.
mains voltage prior to
application of the test
level.
Power frequency
magnetic fields should be
at levels characteristic of
a typical location in a
typical commercial and/or
hospital environment.
Separation distance to
radio communication
equipment must be
maintained according to
the method below.
Interference may occur in
the vicinity of equipment
marked with the symbol:
4-29