GE LOGIQ E9 User Manual page 18

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

Getting Started
Important Notices (continued)
WARNING
CAUTION
CAUTION
1-6
To avoid risk of fire, the system power must be supplied from a
separate, properly rated outlet.
Under no circumstances should the AC power plug be altered,
changed, or adapted to a configuration rated less than
specified. Never use an extension cord or adapter plug.
To help assure grounding reliability, connect to a "hospital
grade" or "hospital only" grounded power outlet.
Use caution to ensure that the power cable does not
disconnect during system use.
If the system is accidentally unplugged, data may be lost.
To avoid leakage current above safety limits as prescribed by
IEC 60601-1 and to ensure continuity of protective earth. DO
NOT connect LOGIQ E9 and mains-operated accessories to a
single or multiple socket extension cord or power strip.
Figure 1-1. Example Plug and Outlet Configurations
1.
100-120 VAC, 1200 VA
Plug and Outlet Configuration
2.
220-240 VAC, 1200 VA
Plug and Outlet Configuration
LOGIQ E9
Direction 5454884-100 English Rev. 1
User Guide

Advertisement

Table of Contents
loading

Table of Contents