Product Overview
1.
Power On/Off Button
2.
Power On Light
3.
Nurse Call Acknowledge/
Volume Control Button
4.
Nurse Call Light
5.
Transmission INOP Light
6.
Battery Low Light
7.
Channel Frequency Label
Chapter 2 - Overview
Telemetry Receiver (Front View)
Press to switch the receiver on.
Off position 0, On position |.
Green LED, lit when the receiver is switched on.
Pressed when Nurse Call activated on the transmitter to
acknowledge the call and stop Nurse Call light flashing
and the intermittent tone sounding. It can also be used
to set Nurse Call volume.
Yellow LED, flashes when the Nurse Call Button is
pressed on the transmitter.
Yellow LED, lit when the transmitter:
•
is switched off.
•
is out-of-range of the receiver
•
is defective.
•
and receiver do not have matching serial numbers
and channel frequency numbers.
•
batteries are exhausted.
Yellow LED, lit when batteries in the transmitter are
low.
Shows the channel number of the receiver. This number
must match the number on the transmitter.